[ Care and Use ManUal ]HSS HPLC Columns1ContentsI. IntRoDUCtIonII. ConeCtInG tHe CoLUMn to tHe HPLC sYsteMIII. WateRs sMaLL PaRtICLe sIze (3.5 �M) CoL
[ Care and Use ManUal ]HSS HPLC Columns2II. CONNECTING THE COLUMN TO THE HPLC SYSTEMa. Column ConnectionHandle the column with care. Do not drop or hi
[ Care and Use ManUal ]HSS HPLC Columns3Figure 4: Waters Ferrule in a Parker Style EndfittingThere are two ways to fix the problem:1. Tighten the sc
[ Care and Use ManUal ]HSS HPLC Columns4e. Measuring Gradient Delay Volume1. Replace the column with a zero dead volume union.2. Prepare eluent A (p
[ Care and Use ManUal ]HSS HPLC Columns5V. CoLUMn InstaLLatIon PRoCeDURe Note: The flow rates given in the procedure below are for a typical 4.6 mm i.
[ Care and Use ManUal ]HSS HPLC Columns6b. Sample Preparation 1. Sample impurities often contribute to column contamination. Use Waters Oasis® or Sep
[ Care and Use ManUal ]HSS HPLC Columns7XIII. CoLUMn CLeanInG, ReGeneRatInG anD stoRaGe a. Cleaning and RegenerationA sudden increase in pressure or s
[ Care and Use ManUal ]Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.comHSS HPLC Columns8